Defibrillator PD 2000

Defibrillator PD 2000
click to enlarge

Manufacturer:
ZOLL

***Excellent condition***
Factory Refurbished unit comes complete with patient cables, 2 batteries, printer, manual, 12 months warranty. In original box.

Ease for Use
A large, bright display provides clear
readable ECG and messages to the
operator during advisory and manual
use. The controls are simple and color
coded for easy operation. Controls are
standardized with other ZOLL products
to reduce training requirements and
provide hospital-wide uniformity of
operation.

Pacemaker
Type: VVI demand; asynchronous (fixed rate)
when used without ECG leads.
Pulse Type: Rectilinear, constant current.
Pulse Duration: 40 milliseconds.
Pulse Amplitude: Variable to 140 mA.
Pacing Rate: Variable from 30 to 180 PPM.
Output Protection: Fully defibrillator-protected and isolated.
Pacer On: Message display on monitor.
Pacing/Defibrillation
Electrodes
Standard configuration allows the use of
ZOLL statpadz or PD 2000 Multi-Function
Electrodes packaged in pairs. Can be
used in the anterior/posterior or anterior/
anterior positions. Sterile and pediatric
electrodes available.
Defibrillator
Waveform: Damped sinusoid.
Output Energy
Delivered: Selectable at 2, 3, 5, 7, 10, 20, 30, 50,
100, 150, 200, 300, 360 joules.
Energy Selection: Control on sternum paddle and front
panel.
Charge Time: Less than 10 seconds.
Delivered Energy
Display: CRT monitor displays delivered energy.
Synchronized
Mode: Synchronizes defibrillator pulse to
patient’s R-wave. “SYNC” message
displayed on monitor. Marker on
monitor and on recorder paper
identifies R-wave.
Charge Controls: Control on apex paddle and on front
panel.
Paddles: Optional paddles are anterior/anterior
(Optional) adult and pediatric. Adult paddles slide
off to expose pediatric paddles.
Defibrillation
Electrodes: Pre-gelled ZOLL PD 2000 Multi-Function
Electrodes, and Multi-Function statpadz
packaged in pairs, can be used in the
anterior/posterior or anterior/anterior
position.
Integral
Defibrillator
Tester: Integral circuitry allows complete test
of defibrillator charge and discharge
without removing paddles from storage
wells. Identical circuitry allows complete
test of unit configured with multifunction
cable.
Defib On: Message display on monitor.
Defibrillation
Advisory: Evaluates electrode connection and
patient ECG to determine if defibrillation
is required.
Shockable Rhythms: Ventricular
fibrillation with amplitude>150μV and
wide QRS complex ventricular
tachycardia with rates greater
than 150 bpm.

General
Size: 10.7 cm high x 33.5 cm wide x 37.4 cm
deep (4.2 x 13.2 x 14.7 in).
Weight: 5.9 kg (13 Ibs) with PD 4400 multi-function
cable; 7.3 kg (16 Ibs) with Power-
Charger attached.
Design Standards: Meets or exceeds all AAMI specifications
for defibrillators. Meets or exceeds UL
544 and CSA standards for medical
equipment and MIL-STD-810E, Method
507.3, MIL-STD 810D Method 514.3,
Section I-3.2.3, Category 3, MIL-STD-
810E, Method 501.3, MIL-STD-810D,
Method 500.2, MIL-STD-202F, Method
213, Condition C, MIL-STD-810E, minimum
integrity test for helicopters.
Patient Safety: All patient connections are fully isolated.
Temperature: Operating: 0°C to 55°C.
Storage: –40°C to 75°C.
Humidity: 5% to 95% relative.